Test/Evaluation Support Equipment

Automated Flow Cell Sorting and Analysis System
Automated Flow Cell Sorting and Analysis System
Bio/cell-based Product Group
Charge : Bio-Medical
Borrowing amount : 50,000 won/hour
Maker : Miltenyi Biotec
Model : AutoMACS Pro Separator
Voucher support status : Yes
  • Available

Principles and Features

* The system uses microbeads-labeled antibodies made from iron oxides and polysaccharides. It is capable of selectively separating cells and is equipped with a robotic arm. Up to 6 samples can be automatically separated.
* Multiple fluorescent proteins such as mCherry, DsRed, GFP, and CFP can be analyzed at the same time. Due to low expression levels, analysis can be performed more sensitively even in the case of dark PE.

Main Specifications

1.Flow Cytometry Main Body
1)Laser: Three air-cooled laser(EXC: 405nm, 488nm, and 561nm)
2)Parameter: FSC, SSC, and eight fluorescence channels
3)Compensation: 7 x 7 matrix (during or post-acquisition), auto-compensation
4)Analysis Rate: Up to 10,000 events/second(up to 106 events/second with the Enrichment Unit)
5)Sampler for automated multi-sample processing(up to 96 samples)

Application Field

It is used when culturing stem cells to collect a single type of cells desired using a specific surface marker that is appropriate from a mix of undifferentiated and differentiated cells or a mix of cells of various origins during the development stage or an immune response, and diverse cells (human, mouse, etc.) are identified and isolated to be used in the field nano biotechnology including immunology, cell biology, molecular cytology.